Braun Ambulances – Damaged Wires and Oxygen Tube May Cause Fire

Braun Ambulances is recalling certain 2020 Demers Ambulances MXP153 Ford F-Series Gas, MXP153 Dodge Ram, and 2020-2021 MXP153 Ford F-Series Diesel emergency vehicles. The medical oxygen tube and electrical wires may become damaged due to insufficient clearance from the ceiling spacers.


Report Received Date 09/23/2021
NHTSA ID 21V740000
Consequence Summary Damage to the wires and medical oxygen tube increases the risk of a fire.
Corrective Action Dealers will remove the ceiling panels and spacers, replace wires and oxygen network tubing, and create a routing channel inside the insulation board for all wires, free of charge. Owner letters were mailed on October 6, 2021. Owners may contact Braun Ambulances' customer service at 1-877-344-9990.
Recall Code NR (Not Reported)
Potentially Affected 9
Fire Risk When Parked No
Do Not Drive No
Recall Link Go to Recall (https://www.nhtsa.gov/recalls?nhtsaId=21V740000)